  • The device "Phase-8"

    Purpose

    Apparatus "PHASE-8" - modern pnevmoelektronnyy device with electric and microprocessor control designed for prolonged ventilator (mechanical ventilation) in adults and children over 6 years in conditions of intensive care and intensive care units fixed medical facilities. In the device provides the possibility of inhalation anesthesia with nitrous oxide. Certificate of 26.12.2002, № ON 1.003.89806-02. The device is introduced into the State Register of medical devices permitted for use in medical practice in Ukraine, № 1597/2003.>

    apparatus equipped with sound and light emergency-warning alarm that is triggered in cases of respiratory gas temperature in excess of over 40 `C, unauthorized disconnection of the power supply, depressurization of the respiratory circuit.

    Technical harakteristika

    adjustment range of minute ventilation in automatic mode at Ti / Tc = 33%, l / min adjustment range of minute ventilation in automatic mode at Ti / Tc = 33%, l / min from 3 to 30

    total minute ventilation range, l / min from 1 to 50

    Control range of respiratory rate, min-1 from 10 to 200

    duration inhalation in the respiratory cycle, from 20% to 80 (from 1:4 to 4:1 TVD / Tvyd)

    inhalation volume, cm3 (l)

    at Ti / Tc = 33%, f = 12 min-1 from 250 to 2500 (from 0.25 to 2.5)

    at Ti / Tc = 33%, f = 200 min-1 from 15 to 150 (from 0.015 to 0.15)

    Switch respiratory cycle time, pressure on

    passive exhalation

    positive end expiratory pressure, cm of water. century. 0 to 20 Maximum inspiratory pressure

    limited safety valve, inches water column 60

    Working pressure inhalation, which installed in the breathing circuit, cm of water. century. (KPa) 10 to 60 (from 1 to 6)

    frequency modes in the forced breaths using a trigger, min-1 from 1 to 29

    frequency modes in the forced breaths using a trigger, min-1 from 1 to 29

    Starts vacuum apparatus from one-to-15 Relative humidity

    respiratory mixture, from 80% to 100

    respiratory gas temperature from 32 to 38

    Range dosed feed gases

    - Oxygen l / min from 0.2 to 10

    - nitrous oxide, l / min from 1 to 10

    emergency oxygen supply, l / min at least 45

    Availability locking device for nitrous oxide is

    range input pressure of oxygen, nitrous oxide, kgf/cm2 from 2 to 6

    power system from AC 220 V, 50 Hz

    Power:

    in the modes of mechanical ventilation, no more than 400 VA

    in the mode of disinfection, not more than 850 VA

    weight apparatus, not more than 55 kg

    Overall dimensions 620x620x1450